Salt for salt spray testing

 
  • Highest purity salt for serious testing
  • For all salt spray testing including the stringent ASTM B117
  • Available in 25 kg drums or bags
  • 25 kg drums are supplied with a scoop to maintain purity when handling
  • Supplied by the leading manufacturer of corrosion test chambers

Specifications

Despite the fact that salt is a commonly occurring compound, pure sodium chloride (NaCl) is much more difficult to come by. Because of its hygroscopic nature, which tends to make salt coagulate and solidify, most granulated salt has anti-caking agents added. However, the addition of these agents is prohibited by most salt spray test standards.

For example, the world's most popular salt spray test standard; ASTM B117, prohibits anti-caking agents, and also limits other impurities, as follows: total impurities < 0.3%, halides excluding chloride < 0.1%, copper < 0.3ppm - so the salt must be very pure indeed, with the limit for copper in particular set at an extremely low level. Other national and international test standards for corrosion testing are similarly demanding in their specifications.

To ensure compliance with the highest standards, Ascott has had Corro-Salt independently analysed by an UKAS* accredited laboratory, who use a combination of X-ray Fluorescence (XRF), Atomic Absorption Spectroscopy (AAS) and Fourier Transform Infrared (FTIR) to verify its purity.

*UKAS = United Kingdom Accreditation Service.

Each consignment of Corro-Salt purchased is supplied with an Ascott certificate of conformity and a copy of the UKAS laboratory analysis. Corro-Salt users can therefore rest assured that they will be testing in full compliance with the requirements of the most stringent standards.

Download the Corro-Salt brochure/datasheet.

Typical Analysis of Corro-Salt

Analysis by X-ray fluorescence 
Total impurities0.145 %
Bromide0.008 %
Fluoridenone detected
Iodidenone detected
Analysis by atomic absorption 
Copper0.1 ppm
Analysis by Fourier transform infrared 
Organicstrace (< 1 ppm)